Instrument technology for UPV/High School (nurse)
From WikiLectures
Fan[edit | edit source]
- A technical device that fully or partially ensures the exchange of gases between the alveoli and the external environment
- The mixture is driven by the so-called supraatmospheric pressure (higher pressure than in the vicinity of the device)
Fan distribution[edit | edit source]
- Distribution by patient age
- → for a neonatal, children's ventilator.
- → universal - enables UPV to be performed in both children and adults.
- → ventilator for adults, where we also take into account the weight of the patient and the size of the cannula.
- Distribution according to use
- →transport ventilator, portable, intended for patient transfer (e.g. Oxylog 3000 from the company Drager or oxygenator).
- → ventilators for performing domestic UPV.
- → anesthetic devices enabling the administration of inhalation anesthetics.
- → ICU ventilators, so-called servo ventilators, the ventilator is also able to monitor the parameters that are measured in the circuit.
- According to the design of the control unit:
- → electronic.
- → microprocessor.
- → mechanical.
- According to the provision of the inspiratory phase
- → pressure, flow, time, volume generators.
- Standard X high frequency.
- Semi-automatic X automatic.
Principle of operation[edit | edit source]
- For a ventilated patient, the inspiratory pressure is generated only by the device (i.e. Ppres) or it is created in combination with the pressure exerted by the respiratory muscles (i.e. Pmus).
- The amount of pressure required for inhalation is given by the sum
- The pressure required to overcome the resistance of the respiratory system (breathing circuit, lung tissue and chest wall).
- The pressure required to keep the DC open.
- The pressure required to overcome the end-expiratory alveolar pressure.
Construction - general scheme[edit | edit source]
- Drive source – compressed gas, electric power.
- Propulsion device – the conversion of energy from a source into the working pressure of a device that performs work.
- Control unit (CTU) – coordinates activities and controls the drive equipment, the fan generation is determined according to the design; hardware and software → generation of devices.
- I. Generation – only mechanical CT (Chirolog 1, Dräger Oxylog 1000).
- II. Generation – ŘJ partially electronic, there are already simple alarms (modern anesthesia devices, Dräger Oxylog 2000).
- III. Generation – a microprocessor enabling electronic feedback and regulation of control valves based on sensed data (Dräger Evita2, Puritan Bennett 7200).
- IV. Generation – multiprocessor ventilators enabling hybrid modes (Dräger Evita XL, high frequency ventilators).
- Control variables include pressure, volume, flow, time.
- Phase variables include trigger, limitation and cycling.
- Expiria modulation device - PEEP valve.
- Control elements – computer screen with a cursor or touch screen or using mechanical control elements (lever, rotary knobs, etc.).
- Pressure and flow measurement:
- The pressure transducer is located either on the Y coupling or in one of the arms of the circuit, by measuring the pressure the function of the device is checked, its activity is monitored, alarms are set, a breath can be initiated according to the pressure gradient.
- The volume is calculated according to pressure and flow over time, the device is called a pneumotachograph or anemometer
- Monitoring unit – an alarm system ensuring safety, displays current values and their trends, triggers an alarm when the limit is exceeded.
- Safety features – backup power supply, backup ventilation mode, anti-asphyxiation valve, leakage compensation, etc.
- The design and layout of the ventilators vary, but regardless, they must meet the basic requirements for setting optimal ventilation for specific patients.
Fan circuit[edit | edit source]
- One-way X two-way system.
- Disposable circuits X reusable circuits.
- Heated circuits X circuits without heating X circuits with double wall hoses.
Basic parameters[edit | edit source]
- FiO2 = fraction of oxygen – percentage of O2 in the breathing mixture (21-100% = 0.21 - 1)
- MV = minute volume – amount of mixture inhaled in 1 min. (MV=Vt xf).
- Vt = volume inhaled in 1 breath (approx. 500 ml).
- PEEP = positive pressure value (in cm H2O).
- P-peak = airway pressure.
- Total frequency: spontaneous + "artificial" breaths.
- Type of ventilation.
Means of humidification and heating[edit | edit source]
- Active humidification – the mixture flows through the chamber system, where it is heated and moistened with sterile water.
- Passive humidification – a moisture and heat exchanger (HME filter) is included in the circuit.
Nebulizers[edit | edit source]
- Jet X Ultrasonic.
- The nebulizer can be connected either to the compressed gas end of the ventilator (aerosol application in synchronization with inspirium) or we include a nebulizer with a continuous flow from the oxygen distribution.
- Spacers.
Monitoring[edit | edit source]
- Oxymeters – used to measure oxygen saturation (0-100% SPo2) and pulse rate (20-250 beats/min.).
- Capnometers – used to measure exhaled CO2.
- Ventilometers – monitor ventilation parameters in the anesthesia equipment circuit (VENAR, ANEMAT, N8).
Suction device[edit | edit source]
- Central - central distribution.
- Mobile - mechanical X electric.
ECMO[edit | edit source]
- Extracorporeal membrane oxygenation used in cases where the lungs are unable to fulfill their function.
- It is a system similar to extracorporeal circulation, where blood is taken from the circulatory system using a catheter, which is then driven through an oxygenator and pumped back into the body.
- Continuous heparinization required.
Veno-venous[edit | edit source]
- Oxygenation of venous blood returned to the venous system.
- Gas exchange only, not circulatory support.
- The drainage cannula is in the right vena jugularis interna → blood is drawn into the pump → oxygenator → return to the circulation in the vena femoralis.
Veno-arterial[edit | edit source]
- Gas exchange and circulation support.
- Drainage cannula in the right internal jugular vein → blood drawn into the pump → oxygenator → return to the patient's circulation via the right common carotid artery.
Links[edit | edit source]
External links[edit | edit source]
References[edit | edit source]
- DOSTÁL, Pavel. Fundamentals of artificial pulmonary ventilation. 2. edition. Maxdorf, 2005. ISBN 80-7345-059-3.