Defibrillation

From WikiLectures

Portable manual external defibrillator.

Defibrillation is an electrical therapeutic method used to reverse malignant heart arrhythmias (ventricular fibrillation, ventricular flutter and hemodynamically significant persistent ventricular tachycardia with unconsciousness) which would inevitably lead to death without intervention. The principle is to pass an electric shock to the patient's myocardium, which causes depolarization of all its fibers, after which the sinus rhythm should be restored. [1] In the case of ventricular fibrillation (VF) we speak of defibrillation, in the case of ventricular tachycardia (VT) we synchronize cardioversion.

Note: electric shock is also used to version some arrhythmias that do not acutely endanger the patient's life (eg atrial fibrillation).

History[edit | edit source]

The study of fibrillation and defibrillation began to develop in the 19th century. It was then found that fibrillation could be induced by the action of alternating current on the heart. In 1899, physiologists Prevost and Batelli discovered that a stronger DC electric shock would stop fibrillation, and then a sinus rhythm appeared. Unfortunately, this discovery did not receive enough attention, but was later confirmed and investigated in many countries.

The first life-saving defibrillation was performed by thoracic surgeon Claud S. Beck in 1947. Thanks to this success, this method was adopted and the foundations of clinical defibrillation research were laid.

Discharge parameters[edit | edit source]

File:Monofazicka sinus.png
Monophasic damped sine wave
File:Monofazicka exp.png
Monophasic cut exponential wave
File:Bifazicka.png
Biphasic wave

Defibrillation requires enough electrical energy to produce the expected effect. If the supplied energy is not sufficient, the discharge must be repeated. Ideal is energy that reverses fibrillation with minimal myocardial damage. In practice, two main types of electric discharges are used:

  • Monophasic devices use a unipolar current either in the form of a damped sinusoidal wave that gradually decreases to zero (more frequent), or a truncated exponential wave that is abruptly terminated before reaching zero.
  • In contrast, in biphasic devices, current flows in one direction for a specified period of time, then reverses and flows in the opposite direction for the remainder of the period. The biphasic device achieves a similar efficiency compared to the monophasic device using approximately half the energy. Therefore, its use is preferred. Each instrument should be marked with a wave characteristic.

Discharge energy is given in joules (E [J] = U [hundreds of volts] · I [tens of amperes] · t [units up to tens of milliseconds]) [2]. According to current recommendations, the first discharge should have an energy of 360 J when using monophasic devices and 150 J in the case of biphasic devices [3].

Types of defibrillators[edit | edit source]

Defibrillators are manufactured in the form of

  • manual direct defibrillators used in open heart surgery
  • manual external defibrillators, which have their place in medical facilities and ambulances,
  • automated external defibrillators (AED) for trained lay people
  • and Implantable Cardioverter-Defibrillator, which are implanted subcutaneously in at-risk patients.

Some devices combine the features of manual and automated defibrillators, allowing external pacing, monitoring ECG, saturation, and the like.

Manual External Defibrillators[edit | edit source]

Manual external defibrillators are used primarily in hospitals, outpatient clinics, and emergency medical service trips. Thanks to the battery power supply, portable versions can be used to advantage (see picture above). Their use is reserved for physicians who decide on the appropriateness of defibrillation.

They consist of a base and a pair of electrodes (slang for "paddles", usually marked according to the target position "sternum" and "apex"). The base has controls for setting the required energy in joules, charging and delivering the shock. The electrode handles can have power selection controls and discharge buttons. Electrodes tend to be reduced to a pediatric size.

Prior to use, the electrodes are coated with a suitable gel or gel sheets are inserted between the electrodes and the patient's skin. When applying the gel, care must be taken not to spread the gel between the electrodes and, in the case of plates, to overlap them, which would result in the discharge passing through the gel layer, not the myocardium, and defibrillation failure. The reduction of the transient resistance between the electrode and the patient's skin is done both by the use of gel or gel plates and by the application of pressure on the electrodes, which should correspond to about 10-15 kilograms. This minimizes the risk of burns.

Automated External Defibrillators (AEDs)[edit | edit source]

Automated External Defibrillator (AED) for lay use.

Approximately 40% of all cardiac arrest is accompanied by ventricular fibrillation or tachycardia [3]. The physician called in can often only notice asystole, so the idea of ​​automated defibrillators has arisen where the patient receives a shock at a time when fibrillation has not yet progressed to asystole.

These defibrillators are designed so that every layman can work with them after a short training. The rescuer opens the device and he guides him through the whole action with instructions on the display or by voice. It has a programmed algorithm according to which rescuers advise what to do. The rescuer sticks electrodes to the patient's chest, the device analyzes his rhythm and recommends or does not recommend the delivery of a shock. The instruments have very high sensitivity and specificity.

AEDs are deployed in places with a high density of people (shopping malls, airports) and where medical assistance is not immediately available (aircraft). They are also located in the homes of at-risk patients. In some countries, defibrillators are also available on the street.

Implantable cardioverters-defibrillators[edit | edit source]

Implantable cardioverter-defibrillator.

These devices are operated on patients at risk of sudden cardiac death in a manner similar to pacemakers. They constantly monitor the patient's heart rate and send a shock if they detect a malignant arrhythmia. During this, the patient feels a certain discomfort. The devices register these events and it is possible to study them later on a computer.

Performing defibrillation[edit | edit source]

Defibrillation can be performed either 'directly' (during open heart surgery or implantable defibrillators), or indirectly (with electrodes applied to the chest surface using higher energy). The electrodes are placed on the chest in standard positions: most often to the right under the right clavicle and to the left to approximately the 5th intercostal space in the middle axillary line. [3] Before use, the electrodes are coated with a suitable gel or (this does not apply in the case of adhesive electrodes). No one should touch the patient during the shock due to the risk of electric shock. During electrical cardioversion, the shock is synchronized with the R-wave according to ECG.

Links[edit | edit source]

Related Articles[edit | edit source]

External links[edit | edit source]

Source[edit | edit source]

  1. MARCIÁN, Paul – CLEMENT, Bronislav – KLEMENTOVÁ, Olga. Electrical cardioversion and defibrillation. Interv Akut Kardiol [online]2011, y. 10, vol. 1, p. 24-29, Available from <http://www.iakardiologie.cz/pdfs/kar/2011/01/05.pdf>. ISSN 1803-5302. 
  2. http://www.resuscitationcentral.com/defibrillation/biphasic-waveform/
  3. a b c The European Resuscitation Council Guidelines for Resuscitation 2005

Used literature[edit | edit source]